MD-50IIA medical equipment 50mA X-ray machine
Main purpose and features:
This X-ray apparatus features fullwave rectification and modular X-ray tube head. The frame of the unit, of cantilever construction, is compact in design easy of movement. The machine can be used for radiography in hospital wards or operating rooms.
Main Technical Data :
Maximum ratings
S kv m mA | 0.08-1 | 2-2.5 | 3.2-4.0 | 5.0-6.3 |
30 | 90 | 90 | 90 | 90 |
40 | 90 | 90 | 90 | / |
50 | 90 | 80 | / | / |
1.Power requirements
Voltage:180-240V Frequency:50Hz
Internal resistance: ≤1ΩRating:5k VA
Current:Maimum15A for radiography
2.Time:0.08 to 6.32 steps electronic
3.Maximum height from X-ray tube focus to floor>1820mm
4.Minimum height from X-ray tube focus to floor<550mm
5.Collimator:Maximum film size at 1000mm
Focal distance: 430×430mm
6.Maximumermote control distance:6m
7.X-ray tube focus:2.3×2.3
8.Unit moving force: <5kg
9.Net weight of the unit:90kg