VET1800B-Veterinary DR
Product Configuration List:
Name | Brand | Quantity | Remark |
X ray High Frequency High Voltage Generator | FSQ30V | 1 set | PLD |
X ray Tube | E7239X | 1 set | Canon |
Collimator | XSQ30 | 1 set | PLD |
Radiography table | SYC60 | 1 set | PLD |
Digital Flay Panel Detector | CareView 1800CW | 1 set | CareView |
DR Acquisition Workstation | DRA40 | 1 set | PLD |
Complete machine accessories | VET1800B | 1 set | PLD |
Production Parameters:
VET1800B | Content | Technical file |
Power Supply | Voltage | 220V±22V |
Frequency | 50Hz±1Hz | |
Capacity | 5kVA | |
Internal resistance | 0.4Ω | |
High Frequency and High Voltage Generator | Power output | 20kW |
Inverter frequency | 500kHz | |
Radiography tube voltage | 40kV-125kV,Step1KV | |
Radiography tube current | 10mA-250mA | |
Radiography time | 1ms-10000ms | |
Control interface | Integrated in image workstation | |
Radiography mAs | 0.2~100mAs | |
Capacitor charging and discharging time | 20S | |
X ray tube assembly(E7239X) | Output power | Big focus 42.5kW small focus 21kW |
Anode heat capacity | 100kJ(140KHU) | |
Maximum heat dissipation of anode | 475W(667HU/s) | |
Component heat capacity | 900kJ(1250KHU) | |
Rotating anode speed | Minimum 2700min-1 | |
Tube Focus: Big Focus/Small Focus | 2.0/1.0 | |
Collimator device (XSQ30) | Collimator light | Halogen lamp,24V/100W |
Visible light illumination brightness | Average illuminance:>100Lux | |
Power supply | 24VAC±1.2V/6.5A | |
Weight | 7.5KG | |
Radiography table | Longitudinal moving range of bed surface | ≧ 500mm |
Bed transverse movement range | Fixed | |
SID | 1000mm | |
Moving Range of radiography Box | Fixed | |
Longitudinal Moving Range pillar | Fixed | |
Table height from the ground | 700mm | |
Table size | 1410mm×670mm | |
Workstation KZT10 | No case,KZT10B | |
Flat Panel Detector(CareView 1800L) | Image sensor | Cesium iodide |
Pixel size | 154μm | |
Effective pixel size | 2816*2816 mm | |
Effective area (H x V) | 434*434mm | |
Image transmission | Gigabit Ethernet (GbitE) | |
Gray scale | 16bit | |
Spatial resolution | 3.4Lp/mm | |
Cycle time | 4-5s | |
Energy range | 40kV to 150kV | |
Power input | AC 220V 50HZ | |
DR Collection workstation DRA40 | Basic function | Newly build patient; newly build examination; anonymous registration; WorkList retrieval and advanced retrieval; WorkList refresh; WorkList information import; checklist; view image; film printing; writing report. |
Collection function | Window width and position adjustment; 90 ° left rotation; 90 ° right rotation; horizontal mirror; vertical mirror; image negative; clipping box; exposure parameter setting; bulb focus switch; exposure mode selection; ionization chamber selection; exposure actual dose value; system status indicator. | |
Image viewing function | Image zoom; local zoom; original size; Restore window size; image grid display; 90 ° left rotation; 90 ° right rotation; horizontal mirror; text mark; line mark; angle mark; arrow mark; clipping box; rectangle mark; circle mark; polygon mark; length correction; four corner information display; ruler display; window width and window position adjustment; sequence image broadcast Put; image output; silhouette function; delete image; image enhancement; reload image; image negative. | |
Film printing function | Add image; insert image; delete image; clear image; image layout; image selection; print format; film print settings; film format settings; add, modify, delete printer settings. | |
Report editing function | Open report; save report; Export Report; add report; delete report; print preview; printer settings; report template editing; image list; add report template; delete case template. | |
Service function | Daily settings: system settings; check protocol editing; system log settings; load user configuration. Advanced settings: restart system; IR settings; IR correction; problem finding; hardware settings; backup recovery; language settings. |