Ø Optional Components
H.F. 80kW Generator: CPI Indio100 80kW Indio100 80kW Technical Specification | |
Application | 80kW Radiographic / RF |
Generator Type | High Frequency 200kHz |
Radiography kVp Range | 40-150kV |
Fluoroscopy kVp Range | 40-125kV |
Radiography mA Range | 10-800mA |
Fluoroscopy mA Range | 0.5-6mA |
mAs | 0.1-1000mAs |
Exposure Time | 1ms-6.3s |
APR, AEC | Support |
Power Supply | 380VAC±38VAC, 3-phase |
Power Frequency | 50/60Hz±1Hz |
Power Capacity | 105kVA |
Internal Resistance of Power Supply | ≤0.11Ω |
Protective Grounding Resistance | ≤4Ω |
X-ray Tubes: Toshiba E7252X Toshiba E7252X Technical Specification | |
Rated Voltage | 150kV |
Rated Focal Spot Value | 0.6/1.2mm |
Anode Input Power | 27/75kW |
Optical anode Angle | 12° |
Anode Heat Storage Capacity | 210KJ (300KHU) |
Heat Storage of Tube Assembly | 900KJ (1250KHU) |
Anode Speed at 50/60/180Hz | 2700/3200/9700rpm |
Inherent Filtration | 0.9 mm Al / 75 kV |
Cooling Method | Natural Cooling |
Weight (without additional components) | 18kg |
ZC25SY-2 Diagnostic Table | |
Table Tilting Range | +90° ~ -25° at 4.4°/s |
Table Length | 2100mm |
Table Width | 625mm |
Movement of Spot Film device | 740mm±10mm( 0° ~ 90° ); 340mm±10mm( 0° ~ 25° ) |
FFD | Changeable 110cm & 150cm |
Transfer Movement of Table Top | 220mm±10mm |
Collimator | Manual/Automatic |
Grid | Fixed, N40, r10, f120cm |
Grid Size | 388×464mm |
Invaray TX-DSI2000 TX-DSI2000 Technical Specification | ||
CCD CAMERA | ||
Resolution | 1K×1K×12bits (1,000,000 pixels) | |
IMAGING INTENSIFIER | ||
Model | Thales TH59438 | |
Fields | 9 inch, 3 fields, 9”/6”/4.5” | |
Center Area Resolution | 54/62/70 lp/cm | |
DQE | 65% | |
MONITOR | ||
Monitor Size | 19" Medical LCD Monitor (1 set) | |
WORKSTATION HARDWARE | ||
PC Computer | Intel Core2 DUO(≥2.3GHz) | |
1G Memory / ≥80G Hard Disk | ||
DVD-RW | ||
TX-DSI2000 SOFTWARE | ||
Operating System | Windows XP | |
Imaging Processing Functions | Digital Fluoroscopy Acquisition 1K×1K/30FPS Digital Spot Acquisition 1K×1K/15FPS Setting window width/level Edge Enhancement Real-time Noise Reduction, Digital Subtract Grey Level Adjustment Zooming and range Automatic/Manuel Electric Cut Image Rotation and Mirror Image Image Cut and Export ShotSave and Restoring image Static high resolution display 4 / 16 frames display on same screen Pixel Displacement and automatic calibration Multiple Image display in single screen Image measurement, measure tool Image notations Patient management Tag tool DICOM print tool Image reset |