MD-8000C Dynamic FPD DRF

Products > DR > MD-8000C Dynamic FPD DRF

MD-8000C Dynamic FPD DRF

Product Use:

This machine is equipped with digital fluoroscopy, digital spot photography and digital DR photography. It can be used for chest and abdominal fluoroscopy, gastrointestinal angiography, urogenital angiography, cholangiography and lower extremity veins. It can be used for head and chest. DR photography of the abdomen and limbs; can also be used for fracture reconstruction under fluoroscopy, foreign bodies, etc.; can be used for peripheral angiography and interventional therapy.


Configuration

Item

Model

Quantity

Remark

Console

KZT70

1

MD-

High voltage generator

FSQ65

1

MD-

X ray tube

E7254FX

1

Canon

Collimator

XSQ40

1

MD-

Diagnostic table

ZDC50

1

MD-

Grid

1.0/1.8m

2

JPI

AEC

SSMC601

1

Clemmond

DAP

d147

1

MD-

Dynamic flat panel detector

DRF-1717RF

1

MD-

Image acquisition workstation

DRA80

1

MD-

Stitching function

optional

1

MD-

Accessory

MD-8000C

1

MD-

  

Technical data:

Item

Content

Technical parameter

Power Supply

Voltage

380V±38V

frequency

50Hz±1Hz

capacity

105kVA

internal resistance

≤0.11Ω

High voltage generator

FSQ65

Maximum output power

65.5kW

Main inverter frequency

440khz, tolerance ± 10%

Radiography tube voltage

40kV~150kV

Radiography tube current

10mA~800mA regulation in steps

Radiography time

1ms~ 10000ms regulation in steps

Radiography mAs

0.1~1000 mAs

Fluoroscopy voltage

40kV~ 125kV Continuously adjustable

 Fluoroscopy   current

0.5mA~ 10mA Continuously   fluoroscopy)

5mA~ 20mA pulse fluoroscopy

Collimator(XSQ40)

Equivalent total   filtration

1mmAL

X ray tube

(E7254FX)

Nominal   anode input power

Big focus   102kW   small focus 40kW

Anode heat   capacity

285KJ(400kHU)  

Component   heat capacity

950kJ(1339kHU)  

Rotating   anode speed

9700rpm180 Hz

Tube focus

Big focus   1.2mm / Big focus 0.6mm

Target angle

12°

Diagnostic

ZDC50

Table   rotating range

+90 ° ~ 0 °   ~ -25 °

Longitudinal   movement of point device

900mm

Table   Lifting

300mm

Lateral   movement range of table surface

220mm

SID

800mm1800mm

Filter grid

18 "×   18" 215lp1 12:1 100cm and 180cm filter grids, one for each, manual   switching

Dynamic flat panel detector     DRF-1717RF

Effective   area

434mm(H)×434mm(V)

Prime matrix

2816(H)×2816(V)

Prime   particle spacing

154μm

Pulse   fluoroscopy

12fps / 1408 x 1408

16fps / 1024 x 1024

22fps /   768x768

Continuous   fluoroscopy

13fps / 1408 x 1408

20fps / 1024 x 1024

30fps /   768x768        

Spatial   resolution

≥3.7lp/mm

A / D Transformation

16bit

Energy range

40 ~ 150 kVp

Working mode

8

Image output   and control

Gigabit lan

Image acquisition workstationDRA80

Workstation software

Registration: routine registration, emergency   registration, adding agreement, adding item, clearing information, starting   inspection;

Work list: list information, patient search to be examined,   refresh to be examined list, delete examination, display column settings.   Start inspection and emergency registration;

Check list: list information, checked patient   display and search, delete image, image storage, disc burning, add item,   display column setting, modify check information;

Patient's body type: thin adult, adult, fat adult;

Photography parameter setting: exposure mode, frame   rate setting, kVp, Ma, MS, MAS, AEC, focus selection;

Perspective parameter setting: exposure mode, frame   rate setting, kVp, Ma, ABS, time reset;

Browsing tools: zoom, horizontal flip, vertical   flip, left turn 90 degrees, right turn 90 degrees, zoom in, zoom out,   original size, moving image, reverse color, adaptive size, ROI magnifying   glass, magnifying glass, default window width and position, window width and   position of interest area, visual window width and position, point gray   value, advanced processing, elliptical gray measurement;

System tools: text mark, front position mark, left   mark, right mark, circle clipping, delete image, delete tool;

Error reset, exposure indication, full screen, save   current image, print

Browsing tools: zoom, horizontal flip, vertical   flip, left turn 90 degrees, right turn 90 degrees, zoom in, zoom out,   original size, moving image, reverse color, adaptive size, ROI magnifying   glass, magnifying glass, default window width and position, window width and   position of interest area, visual window width and position, point gray   value, advanced processing, elliptical gray measurement, image splicing;

Measurement tools: arrow, cardiothoracic ratio   (CTR), distance measurement, angle measurement, spine measurement;

System tools: text mark, front position mark, left   mark, right mark, circle clipping, delete image, delete tool;

Report, save current image, print

Report editing: patient information display and   editing, photo image selection, report content template selection, report   description, report conclusion, report description + conclusion, edit   knowledge base, report doctor, audit doctor, report time, print template,   setting and saving report;

Report printing: fast printing, printing report

Image archiving, burning, printing: delete image,   image storage, browse image, report, lock / unlock, storage queue, print   queue;

Disc recording: volume label, save settings, file   compression, file structure;

Printing: DICOM printer, local printer

System settings: system, annotation information,   tools, others;

Hardware configuration: syncbox, high voltage,   detector, collimator, DAP;

Network configuration: local, worklist, netstore,   local store, printing;

Inspection management: basic information,   positioning information, hardware parameters, image processing, inspection   protocol;

 

Quality management: search, system management;

 

User management: add, update, delete, permission.